1. Click the Windows Start button, then Programs then Cary WinUV and then
Scan.
2. Click the Setup button.
3. Set the following parameters:
Cary tab
X mode: Mode
N
anometers
X mode: Start 1800
X mode: Stop 300
Y mode: Mode % R
Y mode: Y min -5.00
Y mode: Y max 110.00
Show status display ON
Options tab
SBW/Energy: Beam mode Double
SBW/Energy: Slit height Reduced
Source/Detector: Lamps UV-Vis
Source: Source changeover (nm) 350.0
Source: Detector changeover (nm)800.0
Independent tab
Independent control ON
Measurement mode Auto
UV-Vis controls
Ave time (s) 0.1
Data interval (nm) 1.000
Scan rate (nm/min) 600.00
SBW (nm) 2.00
N
IR controls
Ave time (s) 1.000
Data interval (nm) 2.000
Scan rate (nm/min) 120.00
Energy level 3.00
Baseline tab
Correction Zero correction
Autostore tab
File Storage: Storage Storage on (prompt at start)
Note
The NIR detector is noisier than the UV-Vis detector, so a higher Signal
Averaging Time (Ave time) may need to be set in the NIR. To obtain a constant
signal-to-noise ratio over the entire wavelength range, set a value in the Signal-
to-noise field. If you require a faster scan rate, increase the Data interval or
lower the Ave Time.